Product Information

Top Page > Product Information  > Mass flow meters and Controllers > Low-cost Mass Flow SensorMODEL 3810S SERIES

Mass flow meters and Controllers

Low-cost Mass Flow Sensor
MODEL 3810S SERIES
The low-cost mass flow sensor products
The KOFLOC Low-Cost Mass Flow Sensor MODEL 3810S SERIES is the Kofloc's bestseller item. It is a low-cost mass flow sensor designed to function as an internal component of a larger system, with focus on low-cost, accuracy and adaptability.


Download a catalog
Low-cost Mass Flow SensorMODEL 3810S SERIES

KOFLOC MFS MODEL 3810S SERIES

Quick Review

Low-Cost
The KOFLOC Low-Cost Mass Flow Sensor MODEL 3810S SERIES is a time-tested bestseller mass flow sensor intended to function as an internal device. Its low-cost is achieved through exhaustive cost-cutting methods with the final price approaching those of variable area flow meters.

KOFLOC DependabilityKOFLOC has 60 years of experience in manufacturing precision flow instruments. KOFLOC products, both standard and off-catalog custom items are well received throughout Japan as the name that manifests confidence. KOFLOC, the fluid expert, is the name professionals trust in precision control of minute flow.

0 to 5 V Analog Output
The KOFLOC Low-Cost Mass Flow Sensor MODEL 3810S SERIES is a mass flow device. It requires no temperature or pressure dependant compensations or calculations. 0 to 5 V analog output can also be used as a part of recording, control and alarm devices.

Accuracy
The KOFLOC Low-Cost Mass Flow Sensor MODEL 3810S SERIES is a low-cost MFS, but it employs a bypass capillary mechanism and constant current thermal differential sensor that are commonly used in more expensive devices.

TOP


KOFLOC Mass Flow Sensor MODEL 3810S SERIES may be used for the following gases.
Medium (gases): Actual gas calibration is available for the gas listed
Air
Argon Ar
Carbon dioxide CO2
Helium He
Hydrogen H2
Nitrogen N2
Oxygen O2

Medium (gases): N2 equivalent conversion factor (CF) is used for the gas listed. Contact KOFLOC for the availability of actual gas calibration.
Acetylene (Ethyne) C2H2
Ammonia NH3
Arsine AsH3
Boron trichloride BCl3
Boron trifluoride BF3
Butadiene (1,3-) C4H6
Butane C4H10
Butene (1-) C4H8
Butene (2-) (Cis) C4H8
Butene (2-) (Trans) C4H8
Carbonylsulfide COS
Carbon monoxide CO
Chlorine Cl2
Cyanogen C2N2
Cyclopropane C3H6
Deuterium D2
Diborane B2H6
Dichlorosilane SiH2Cl2
Dimethylamine C2H6NH
Dimethylether C2H6O
Disilane Si2H6
Ethane C2H6
Ethylene (Ethene) C2H4
Ethylchloride C2H5Cl
Freon-113 C2Cl3F3
Freon-114 C2Cl2F4
Freon-115 C2ClF5
Freon-116 C2F6
Freon-12 CCl2F2
Freon-13 CClF3
Freon-13B1 CBrF3
Freon-14 CF4
Freon-22 CHClF2
Freon-C318 C4F8
Hydrogen bromide HBr
Hydrogen chloride HCl
Hydrogen cyanide HCN
Hydrogen iodide HI
Iso-Butane C4H10
Isobutylene (Isobutene) C4H8
Krypton Kr
Methane CH4
Methylbromide CH3Br
Methylmercaptan CH3SH
Monomethylamine CH3NH2
Neon Ne
Nitric oxide NO
Nitrous oxide N2O
Ozone O3
Pentane C5H12
Phosgene COCl2
Phosphine PH3
Propane C3H8
Propylene (Propene) C3H6
Silane SiH4
Sulfur dioxide SO2
Sulfur hexafluoride SF6
Trimethylamine C3H9N
Xenon Xe

KOFLOC Mass Flow Sensor MODEL 3810S SERIES may NOT be used for the following gases.
Medium (gases)
Bromine pentafluoride BrF5
Carbon disulfide CS2
Chlorine trifluoride ClF3
Cyanogen chloride ClCN
Dibromo difluoromethane Br2CF2
Ethylene oxide C2H4O
Fluorine F2
Freon-11 CCl3F
Freon-21 CHCl2F
Hydrogen fluoride HF
Hydrogen sulfide H2S
Methylchloride CH3Cl
Nitrogen dioxide NO2
Nitrogen trifluoride NF3
Vinylchloride C2H3Cl
Vinylfluoride C2H3F

Note: If you did not find the right gas, please contact KOFLOC for more information.


TOP

Features

Rohs compliance
Light compact build
Free mounting position
Wide selection of fittings, seals, wetted materials and connections
Easy maintenance

TOP

Applications

Air sampling
Analyzer component
Bioreactor
Biotechnological process
Burn control
CCFL manufacturing
Chamber pressure control
Chemical process
Environment monitoring
Electronic device manufacturing
Food production
Fuel cell testing
Fermentation process
Gas blending
Gas distribution
Gas flow monitoring
Gas generation
Industrial furnace
Leak testing
Pharmaceutical process
Pollution monitoring
Process flow control
R&D
Solar power element manufacturing
Purge gas measurement

TOP

Standard Specifications

Flow range(at N2 calibration conditions)F.S.10~2SLMF.S.3~50SLM
AccuracyWithin±3.0%F.S.Within±5.0%F.S.
RepeatabilityWithin±0.5%F.S.
Proof pressure980kPa
Working temperature range5 to 45°C(Accuracy guarantee:15 to 35°C
Materials of parts in contact w/gasesBody:SUS 303,PTFE
Seals:VitonTM
JointStandard:1/4RC
Flow output signals0~5VDC
Required power supply+15VDC(±5%)40mA, -15VDC(±5%)10mA
Weight250g350g

Cable Connections

No.1Power source +15VDC
No.2Power source COM
No.3Power source -15VDC
No.4Flow output 0~5VDC
No.5Flow output COM
Note: The Series 3810 pin assignment has been modeified from March 2000, and is not compatible with the old model. Please check the following:

AMP171826-5 on the Connector 3810 side
AMP171822-5 on the cable side


TOP

About this commodity

Inquires concerning products The download of various material is here